Reconstructed reference notes on EU and UK medical-device software, built from six sources captured 2026-10-13: MDCG 2019-11 rev.1 (June 2025), MDCG 2020-6 (April 2020), MDCG 2020-13 (July 2020), MDCG 2020-7 (April 2020), MDCG 2020-8 (April 2020), and the MHRA software-applications guidance (2023-07-01) with its stand-alone software PDF. Use for SaMD qualification and classification, legacy-device clinical evidence, the clinical evaluation assessment report, the PMCF plan, and the PMCF evaluation report, under EU MDCG guidance or UK MHRA guidance. SCOPE LIMITS: EU MDR and UK MHRA device software only; no IEC 62304, ISO 13485, or ISO 14971 text; no MDR or IVDR regulation body text; MDCG guidance is not legally binding; MHRA guidance is UK-specific; synthesized notes, not legal advice. LICENCE: Per-document: European Commission reuse policy (Decision 2011/833/EU, CC BY 4.0, attribution and indication of changes) for the MDCG documents; Open Government Licence v3.0 for the MHRA guidance; reconstructed notes with attribution; official PDFs not redistributed.

Overview

Publisherjgsystemsconsulting
Repositoryjgs-se-knowledge-packs
Skill namemed-device-eu
Stars
121
Forks
1
Bundled files
11
Links
  • Markdown instructions

    A SKILL.md file the model loads on demand, so it only costs tokens when a request actually matches.

  • Works with any LLM

    AI skills are plain Markdown, not provider-specific code, so this works with GPT, Claude, Gemini, Grok, or a local model.

  • 11 bundled files

    Scripts, templates, and references the model can read while it works. Files are read-only and never executed.

  • Open source

    Published by jgsystemsconsulting on GitHub. Read the source before you install it.

Installation

Install the Med Device Eu AI skill in TypingMind to use it with any LLM, or drop it into another agent that reads SKILL.md.

1

Install in TypingMind

TypingMind installs a skill straight from its GitHub folder — it reads SKILL.md, bundles the resource files, and stores the result locally.

  1. Open the app and go to Plugins → Skills.
  2. Choose "Install from GitHub".
  3. Paste the skill folder URL below and confirm.
    https://github.com/jgsystemsconsulting/jgs-se-knowledge-packs/tree/main/packs/med-device-eu
  4. Enable the skill in any chat where you want it available.
Plugins → Skills → Add skill → From GitHub URL, then paste the folder URL and press Continue.
2

Install in another agent

Any agent that reads the Agent Skills format can use this skill — copy the folder into that agent's skills directory.

Claude Code — .claude/skills
git clone --depth 1 https://github.com/jgsystemsconsulting/jgs-se-knowledge-packs.git /tmp/jgs-se-knowledge-packs
mkdir -p .claude/skills
cp -r /tmp/jgs-se-knowledge-packs/packs/med-device-eu .claude/skills/med-device-eu
Restart Claude Code after copying so it picks up the new skill.

Use it in TypingMind

Enable Med Device Eu in any TypingMind chat and the model takes it from there. Its name and description sit in the system prompt, and the moment a request matches, the model loads the full instructions itself — you never invoke it by hand, and it costs no tokens until it is actually used.

The model loads Med Device Eu on its own as soon as a request matches it.

Works with any AI model

AI skills are plain Markdown instructions rather than provider-specific code, so Med Device Eu is not tied to the model it was written for. Install it once in TypingMind and use it with GPT-5, Claude, Gemini, Grok, DeepSeek, Mistral, Llama, or a local model you run yourself — all on your own API keys.

  • Loaded only when it is needed

    The system prompt carries just the name and description. The instructions are fetched on the first matching request, so an idle skill costs nothing.

  • Switch models mid-chat

    Because the skill is instructions rather than code, changing model does not break it — the next model reads the same SKILL.md.

Skill instructions

This is the SKILL.md content the model loads. Read it before installing — a skill is instructions your model will follow.

EU and UK Device Software Guidance (MDCG and MHRA)

Sources: MDCG 2019-11 rev.1 (June 2025), MDCG 2020-6 (April 2020), MDCG 2020-13 (July 2020), MDCG 2020-7 (April 2020), MDCG 2020-8 (April 2020), and MHRA Medical devices: software applications (2023-07-01) with its stand-alone software guidance | Licence: per-document, see LICENSE | Chapters: 8

When to use

Prerequisites: none: plain Markdown; no MCP server, API key, or licence tier needed at runtime.

Use this pack for questions about software in and around medical devices on the EU and UK side: whether an app or stand-alone package qualifies as a device or an IVD and which risk class applies, what clinical evidence a legacy device needs under the MDR, how a notified body documents its assessment of a clinical evaluation, and what a post-market clinical follow-up plan commits to and how the evaluation report closes it. Both readings are taught: the MDCG guidance route for the EU and the MHRA guidance route for Great Britain.

How to Use This Skill

  • Without arguments: read the Core Frameworks below, covering the two-sided qualification question, the clinical-evidence standard for legacy devices and its assessment, and the plan-and-report loop for post-market clinical follow-up.
  • With a topic: use the Topic Index to find the chapter, then ask directly (e.g. "is this symptom checker a device in the UK", "what class is this dosing software", "what does the PMCF plan have to promise").
  • With a chapter: ch01 is the landing chapter for "which regime applies"; ch02-ch03 qualification and classification; ch04-ch05 clinical evidence and its assessment; ch06-ch07 PMCF plan and report; ch08 UK obligations after qualification.

Core Frameworks & Mental Models

SaMD qualification: one question, two readings (ch02 MDCG; ch03 and ch08 MHRA)

Qualification asks whether software is a device at all, and this pack teaches both readings. The EU route (ch02) runs the MDCG five-step decision: is it software, does it touch an Annex XVI product or drive or influence a hardware device, does it act on data beyond storage, archival, communication, simple search, or lossless compression, is that action for an individual patient's benefit, and does it carry a medical purpose of its own. Classification then follows Rule 11 of Annex VIII: class IIa for information used in diagnostic or therapeutic decisions, rising to IIb or III as the harm from wrong information rises; IIa to IIb for physiological monitoring; class I for everything else. The UK route (ch03) reads medical purpose from labelling, app-store text, and other claims, keeps a verifiability discretion for calculators the intended user can easily check, and draws the decision-support line by whether software reproduces a paper path or produces a diagnosis, prognosis, or individual risk figure. Once software qualifies on the GB market, the obligations follow (ch08): UKCA marking, a classification route, registration with the MHRA, essential requirements, and post-market duties. The mental model: both readings share an ancestor, so the medical-purpose test is similar; the divergences are in the details, and they bite hardest on calculators and decision support.

Clinical evaluation: legacy evidence and the assessment report (ch04, ch05)

For a device CE marked under the old directives (ch04), sufficient clinical evidence is the present result of a qualified assessment, not a fixed dataset. The directive-era data carries over into a gap analysis against the MDR's narrower clinical-data definition and stricter equivalence rules, and each dataset is appraised for methodological quality and bias. Two rules carry weight: complaint and incident rates alone cannot prove safety, and where the evidence will not carry the claimed intended purpose, the fix is to narrow the purpose. Well-established technologies may run on a lower evidence level; the non-clinical route exists but is exceptional. The notified body's side of the same work is the assessment report (ch05): a per-section template whose verdicts map the manufacturer's clinical evaluation, with special sections for expert-panel consultation and the non-clinical route. The mental model: the manufacturer writes the evidence case, the assessment report grades it, and both stand on the MDR's evaluation stages.

PMS and PMCF: the plan-and-report loop (ch06, ch07)

Post-market clinical follow-up is a continuous process that updates the clinical evaluation, run as a controlled document pair inside post-market surveillance. The plan (ch06) commits every activity to one of five aims, states where the need came from, justifies method, sample size, and endpoints, and fixes timelines at quarterly or yearly resolution; a decision not to run PMCF at all needs its own justification. The report (ch07) closes a named, versioned plan: every performed activity reported, deviations justified, findings discussed whether positive or negative, and every finding routed into the clinical evaluation report, the risk management file, the PMS plan, and the SSCP. The mental model: the plan promises, the report reconciles, and the conclusions feed the next round of all four documents.


Chapter Index

#ChapterKey content
ch01EU and UK orientationQuestion-to-document map; MDR, IVDR, and UK MDR 2002 as the binding layers; guidance as non-binding but operative (S1-S6)
ch02SaMD qualification and classificationThe five-step qualification decision, Rule 11 sub-rules, the IVD MDSW track, rev.1 changes, modular placement (S1)
ch03MHRA software qualificationThe boundary test, the app examples, the calculator verifiability discretion, where the UK reading diverges from the MDCG one (S6)
ch04Clinical evidence for legacy devicesThe sufficient-evidence definition, the gap analysis, appraisal tools, the evidence hierarchy, well-established technologies (S2)
ch05Clinical evaluation assessment reportCEAR sections A to H, per-section verdicts, the special sections for panel consultation and the non-clinical route (S3)
ch06PMCF planThe five aims, per-activity justification, registry and real-world evidence rules, the hard study requirement for implantable and class III devices (S4)
ch07PMCF evaluation reportPlan-bound closure, per-activity reporting, the routing table, the conclusions loop (S5)
ch08MHRA UK device obligationsUKCA marking and placement, classification routes, essential requirements, registration, vigilance, field safety, end of life (S6)

Topic Index

  • SaMD qualification → ch02, ch03
  • software classification → ch02
  • legacy device clinical evidence → ch04
  • clinical evaluation assessment report → ch05
  • PMCF plan → ch06
  • PMCF evaluation report → ch07
  • MHRA software apps → ch03
  • UKCA device software → ch08
  • which regime applies → ch01

Supporting Files

  • glossary.md: the ten working terms (SaMD, legacy device, PMCF, PMS, clinical evaluation, CEAR, MDR, IVDR, UKCA, MDCG), each with the chapter that teaches it
  • cheatsheet.md: decision shortcuts for the qualification question, the template to reach for, and which regime a question belongs to

Scope & Limits

Covers: the EU and UK regulatory treatment of medical device software as stated in the six captured sources: qualification and classification, legacy-device clinical evidence, the clinical evaluation assessment report, the PMCF plan and evaluation report, and UKCA-era obligations for qualified software.

The binding text is the regulation: Regulation (EU) 2017/745 (MDR), Regulation (EU) 2017/746 (IVDR), and the UK Medical Devices Regulations 2002. This pack names them and does not reproduce them. A reader who needs the regulation text reads it on EUR-Lex or legislation.gov.uk.

The US counterpart is fda-med-device. That pack owns FDA questions, including 510(k) and De Novo pathways and FDA risk categorization; this pack does not answer FDA questions.

Scope limits, repeated: IEC 62304, ISO 13485, and ISO 14971 are named only and not reproduced; MDR and IVDR are named only and not reproduced; the pack is synthesized reference notes and not legal advice; MDCG guidance is not legally binding; MHRA guidance is UK-specific. These documents revise, so check for the current revision before relying on any single requirement.

Bundled files

The model reads these on demand while the skill is loaded. They are exposed as readable files and are never executed.

Frequently asked questions

What does the Med Device Eu AI skill do?

Reconstructed reference notes on EU and UK medical-device software, built from six sources captured 2026-10-13: MDCG 2019-11 rev.1 (June 2025), MDCG 2020-6 (April 2020), MDCG 2020-13 (July 2020), MDCG 2020-7 (April 2020), MDCG 2020-8 (April 2020), and the MHRA software-applications guidance (2023-07-01) with its stand-alone software PDF. Use for SaMD qualification and classification, legacy-device clinical evidence, the clinical evaluation assessment report, the PMCF plan, and the PMCF evaluation report, under EU MDCG guidance or UK MHRA guidance. SCOPE LIMITS: EU MDR and UK MHRA device sof...

Why use Med Device Eu on TypingMind?

Because you install it once and use it with any model. Med Device Eu is plain Markdown rather than provider-specific code, so the same skill runs on GPT-5, Claude, Gemini, Grok, or a local model — and you can switch model mid-chat without it breaking. TypingMind runs on your own API keys, so you pay providers directly instead of a per-seat subscription, and your skills and chats stay in your own storage.

How do I install Med Device Eu in TypingMind?

Open Plugins → Skills → Install from GitHub in TypingMind and paste https://github.com/jgsystemsconsulting/jgs-se-knowledge-packs/tree/main/packs/med-device-eu. TypingMind reads its SKILL.md and bundles its files and installs it as a skill you can enable per chat.

Which AI models can use Med Device Eu?

Any model you connect in TypingMind. AI skills are plain Markdown instructions rather than provider-specific code, so GPT, Claude, Gemini, Grok, and local models can all load this skill when a request matches it.

How many AI models can I use with Med Device Eu?

As many as you like. As long as a model supports skills, you can use Med Device Eu with it — GPT, Claude, Gemini, Grok, DeepSeek, Mistral, Llama and more — all on TypingMind with your own API keys.

Is the Med Device Eu AI skill free?

It is published on GitHub by jgsystemsconsulting. Check the repository for licensing terms. You only pay your own AI provider for the tokens you use.

What are AI skills?

An AI skill is a reusable instruction bundle that teaches an AI model how to do one specific task. It follows the open Agent Skills format: a SKILL.md file with a name and description, plus any scripts, templates or reference files the model may need. The model reads the instructions only when your request matches the skill, so an installed skill costs nothing until it is used.

How are AI skills different from plugins or MCP servers?

A plugin or MCP server gives a model new tools to call — code that runs somewhere and returns a result. An AI skill gives the model knowledge and process instead: how to approach a task, which steps to follow, what good output looks like. Skills are plain Markdown, so they need no server, no API key and no runtime, and they work with any model.